NMRFAM SampleJet Policy and Instructions

WARNING: Never exit topspin with your sample in the magnet and the SampleJet in manual mode! Your PI will pay for the damage caused by your sample tube being crashed by the SampleJet inside the (cryo)probe (Probe cost ~$125,000- $150,000)!

Before you do anything with the SampleJet, make sure that “no NMR experiment is still actively being run by someone else”. Therefore, the rule is: Start topspin before you operate the SampleJet to avoid unintentionally ejecting another user’s sample while an experiment is running. If topspin won’t run due to a licensing issue, someone else is most likely still logged in and using the software. Seek advice on how to proceed from this point (see Kaifeng, Gabriel, Mark, or Milo, in whatever order works).

By default, the SampleJet is to be left in Automatic Mode (1.7mm Shuttle mode for Kurgan, 5mm Shuttle mode for Kerry and Rotvieh).

To change to Manual Mode:

Open the web browser (there is a web browser icon on the top panel after system)
Type in samplejet in the address bar (or enter http://149.236.99.55/)
Click Login tab on the left, click Login in the sub-menu
Login as User, with password bruker. System will show Login as User successful.
Click Basic Service, then Basic Commands in the sub-menu.
Go to the section Change Operating Mode, click Manual. This will automatically bring whatever is inside the magnet (the shuttle and sample) back into position within the SampleJet. The system will automatically restore the sample to the appropriate location.
Be patient, watch the status of the SampleJet at the bottom of the webpage until the Operating Mode displays Manual. You can now use the standard eject (ej) and inject (ij) commands to raise and lower the sample and spinner.

Once you are done with your experiments, you are RESPONSIBLE for removing your sample from the magnet before you exit topspin!

To change back to “Automatic Mode”:

Once your sample is removed, you MUST change the SampleJet back to automatic mode (1.7mm Shuttle mode for Kurgan, 5mm Shuttle mode for Kerry and Rotvieh).
Go to the same webpage, read the warning message in red at the bottom Make sure magnet is empty before you leave manual mode!. That is, you must eject your sample first.
Go to the section Change Operating Mode, click 1.7mm-Shuttle for Kurgan, 5mm-Shuttle for Kerry and Rotvieh. Wait for a while, the SampleJet will return to the default automatic mode.

WARNING: Never exit topspin with your sample in the magnet and the SampleJet in manual mode! Your PI will pay for the damage caused by your sample tube being crashed by the SampleJet inside the (cryo)probe (Probe cost ~$125,000- $150,000)!