3D NOESY-13C-1H-HSQC


3D NOESY-13C-1H-HSQC with optional 15N decoupling, 13C' SEDUCE-1 decoupling and 2H decoupling

References

Muhandiram et. al, J. Magn. Reson. B 102, 317-321 (1993)

Acquisition

Pulse program

c13noehsqc.fa

Gradient program

Gradient program: c13noehsqc.r

Include files etc.

Include file for pulsed field gradient definitions: Gradfa.incl

Include file for optional expt time correction: faexptcorr.incl

Composite pulse decoupling program for SEDUCE-1 decoupling: waltz16sp3p31

Shaped pulse file for compensated SEDUCE-1 decoupling: seduce1c6.jc (JCAMP-DX), seduce1c6.rf (binary RF format) or seduce1c6.txt (ASCII) (cosine modulated SEDUCE-1 shape with excitation optimum at 6*(1/P90) Hz from carrier)

Shaped pulse file for compensated SEDUCE-1 decoupling: seduce1c5.jc (JCAMP-DX), seduce1c5.rf (binary RF format) or seduce1c5.txt (ASCII) (cosine modulated SEDUCE-1 shape with excitation optimum at 5*(1/P90) Hz from carrier)

Read more about the pulse shape file formats.

Parameters

Acquisition parameters: 750 MHz

Processing

Use ux2flx or equivalent utility to convert your Bruker data to Felix "dat" format.

Felix95 processing macro: ft3d_states.mac

Also used (for correction of digital filter artifacts): digphs.mac


| Top of Page | Main Index |